Products

Vertical Laminar Flow CabinetBKCB-V1100

Vertical Laminar Flow CabinetBKCB-V1100

Description:

ADVANTAGE

1.The  cabinet  is  designed  with   a  vertical   surface,the  operation  perspective  is  wide   and  more  humanized.

2.Made   of   high   quality   cold    rolled   steel   plate    material,welding   structure,surface   electrostatic   spraying,enhancing the  stability  of  the  cabinet.

3.Equipped  with   initial   effect   filter,which   can   preliminarilyfiter   large   particulate   matter,thus   improving   the   service life  of  the  high  efficiency  filter.

4.Work  surface  is  made  of  high  quality  304  stainless  steel  material,beautiful  and  corrosion  resistance.

5.The   working   area   adopts   four   sides   (left   and   right   sides,rear,bottom)positive   pressure   surround   design,effectively protect   product   safety.

6.The  front  window  adopts  5mm  thick  tempered  glass  to  better  protect  the  safety  of  personnel  and  experiments.

7.The side windows are  made of 5mm thick toughened glass to enhance the  lighting  permeability of the operation area.    8.The  hidden  weight  structure,manual  front  window,realizing  freely  adjusting  the  opening  height  of  the  front  window.

9.The  base  bracket  and  casters  are  integrated  design,which  can  be  moved  safely  through  the   casters and  can  also  be  fixed  and  leveled.

10.7-inch  color  touch  screen,real-time  display  of  running  status   (air  flow   direction,air  velocity,machine   running   time, UV   lamp   running   time,high   efficiency   fiter   running   time,high   efficiency   filter   running   resistance,operating   area

temperature   and   humidity,clock   display)at   a   glance.

11.With  the  function  of  appointment  timing,it  can  automatically  set  the  time  of  starting  and  shutting  down  the  laminar flow  cabinet  and  the  disinfection  time  of  ultraviolet  lamp.

12.With  a  key  disinfection  function,convenient  and  fast,greatly  saving  the  work  time  and  improving  the  work  efficiency. 13.Perfect    alarm    system:

(1)The  front  window  is  not  set  height  alarm:the  front  window  is  set  height  of  200mm,when  the  front  window is  not  in  the  height  position  and  the  bottom  position,sound  alarm  will  be  issued.

(2)Ultra-high  filter   pressure  alarm:when  the  filter  resistance   increases,sound  alarm  will   be   issued.

(3)Filter  failure  replacement  alarm:when  the  fiter  service  life  expires,there  will  be  a  filter  replacement  sound  alarm. 14.Safety    interlock    protection:

(1)Fan  and  front  window  interlock:when  the  front  windowdoor  falls  to  the  bottom,the  fan  will  automatically  close.

(2)The  ultraviolet  lamp,the  front  window   and  the   lamp   interlock:the  front  window   door  falls  to  the   bottom

and  the  lamp  is  not  opened,the  ultraviolet  lamp  can  be  opened  to  avoid  the  misoperation  ofthe  ultraviolet  lamp causing  harm  to  the  human  body.